Thermo TSQ Altis Valve Issue
Hello,
We have Thermo Fisher Altis TSQ II MS. In my previous post I talked about inconsistent data post-preventative maintenance (PM). I have now realized that what I’m calling “data” post-PM are just artifacts with intensity scale 10^0 or 10^1 whereas the real analyte peaks are 10^6. I also just realized that the divert valve to my MS from LC is not connected. See the attached picture from Chromeleon showing valve “NC” which I believe means not connected. Moreover, in the command area, it also reads DivertValve A properties “DV_NOT_CONNECTED”. Subsequently, when I changed instrument method for my sequence to DivertValve A “1-2” (which is our flow to MS), it gave me an error that DivertValve is not connected.
I checked instrument controller/configuration tab but there Divert Valve is set to A so I’m not sure why Chromeleon is not picking it up.
Is this a hardware or software issue? I am already contacting Thermo about it but they’re soo slow. Any help would be appreciated!
EDIT: the field service engineer also updated our Chromeleon to the latest version during PM (our version was like 4 years old).